13.2 Input Specifications/Output Specifications/Measurement Specifications
215
10
13
Cha
p
ter 13 Spe
cificatio
ns
Specified for RMS voltage Urms and RMS current Irms. Current and power values incorporate current sensor accuracy.
-6. Flag concept
IEC61000-4-30 Flagging concept
If an unreliable values are produce during a dip, swell, or interruption, approx. 200-ms aggregation will be
“flagged.”
An interval data including the flagged 200-ms aggregation will also be flagged.
Flagged data are referenced to decide the frequency for an interruption, and are recorded in status informa-
tion of the TIME PLOT data. If events of a dip, swell, or interruption are set to off, the values are also flagged.
(37) Mains signaling voltage Msv, Msv%
Measurement method
Compliant with IEC61000-4-30
Levels (Msv) or content rates compared to the nominal voltage (Msv%) are calculated based on
the mid-harmonic bin of 10/12-cycle RMS values of up to two set signal frequencies or four mid-
harmonic bins that most closely approximate those frequencies to display.
Displayed item
Msv1, Msv%1, Msv2, Msv%2, the worst value between event IN and event OUT
Measurement range
600.00 V
Measurement accuracy
Within the range of 3% to 15% of nominal voltage: ±5% rdg.
Within the range of 1% to 3% of nominal voltage: ±0.15% of nominal voltage
Event threshold
Percentage of the nominal voltage
Event IN
Start time of approx. 200 ms aggregation in which the Msv value exceeds the threshold value
Event OUT
Depends on the set timeout.
Multiple-phase system treatment
Starts when any one of the channels from U1 to U3 exceeds the threshold value.
Saved waveforms
Available
Constraints
Not displayed for 400 Hz measurement.
-5. RMS frequency characteristics
Frequency
Voltage
Current
Power
40 Hz to 70 Hz
Specified as RMS value
Specified as RMS value
Specified as RMS value
70 Hz to 360 Hz
±
1% rdg.
±
0.2% f.s.
±
1% rdg.
±
0.5% f.s.
±
1% rdg.
±
0.5% f.s.
360 Hz to 440 Hz
Specified as RMS value
Specified as RMS value
Specified as RMS value
440 Hz to 5 kHz
±
5% rdg.
±
0.2% f.s.
±
5% rdg.
±
0.5% f.s.
±
5% rdg.
±
1% f.s.
5 kHz to 20 kHz
±
5% rdg.
±
0.2% f.s.
±
5% rdg.
±
0.5% f.s.
±
5% rdg.
±
1% f.s.
20 kHz to 50 kHz
±
20% rdg.
±
0.4% f.s.
±
20% rdg.
±
0.5% f.s.
80 kHz
-3 dB
-3 dB
HIOKI PQ3198A961-03
Summary of Contents for PQ3198
Page 2: ...HIOKI PQ3198A961 03...
Page 6: ...Contents iv HIOKI PQ3198A961 03...
Page 20: ...Usage Notes 14 HIOKI PQ3198A961 03...
Page 44: ...2 3 Display Items and Screen Types 38 HIOKI PQ3198A961 03...
Page 58: ...3 8 Turning the Power On and Off Setting the Default Language 52 HIOKI PQ3198A961 03...
Page 102: ...5 8 Factory Settings 96 HIOKI PQ3198A961 03...
Page 198: ...12 4 Downloading Recorded Data to Computer 192 HIOKI PQ3198A961 03...
Page 264: ...14 4 Disposing of the Instrument 258 HIOKI PQ3198A961 03...
Page 296: ...A32 HIOKI PQ3198A961 03...
Page 299: ...Index Index 3 Index Wiring 73 Z Zero adjustment 53 Zero phase A28 HIOKI PQ3198A961 03...
Page 300: ...Index Index 4 HIOKI PQ3198A961 03...
Page 301: ...HIOKI PQ3198A961 03...
Page 302: ...HIOKI PQ3198A961 03...
Page 303: ...HIOKI PQ3198A961 03...
Page 304: ...www hioki com HIOKI PQ3198A961 03...