38
Measuring Line-to-Line Voltages in a Three-Phase Circuit
Phase voltage measurement (values for reference purposes)
The display will show V
R
, V
S
, and V
T
(or V
1
, V
2
, and V
3
).
When measuring a three-phase circuit, the display will show the
three phase voltage values. However, since the neutral line cannot
be measured, the displayed values indicate a line-to-ground
voltage of each phase using the virtual neutral point (ground) as the
reference.
The displayed phase voltages, accuracy of which is not
guaranteed, can be used for reference only.
1
Display the line-to-line voltage
measurement screen.
The line-to-line voltage measurement screen
is displayed first at the time of the start-up.
Press and hold
2
Press and hold the
Fn
key.
The phase voltage measurement screen will
be displayed.
3
Check the displayed values.
If the phase voltage is less than 30.0 V, the
string
[Lo]
will be shown; if the phase voltage
exceeds 400.0 V, the string
[ovEr]
will be
shown.
4
Press the
Fn
key.
The line-to-line voltage measurement screen
will be displayed.
HIOKI PD3259C961-00
Summary of Contents for PD3259-50
Page 2: ...HIOKI PD3259C961 00...
Page 6: ...iv Contents HIOKI PD3259C961 00...
Page 26: ...20 Part Names and Functions HIOKI PD3259C961 00...
Page 38: ...32 Inspecting the Instrument HIOKI PD3259C961 00...
Page 64: ...58 General Specifications Accessories See p 2 Options See p 3 HIOKI PD3259C961 00...
Page 74: ...68 Troubleshooting HIOKI PD3259C961 00...
Page 78: ...Voltage Sensors of the PD3259 50 72 HIOKI PD3259C961 00...
Page 82: ...Index 76 W Wireless Adapter 3 28 52 Wireless communications 28 52 HIOKI PD3259C961 00...
Page 83: ...HIOKI PD3259C961 00...
Page 84: ...HIOKI PD3259C961 00...
Page 85: ...HIOKI PD3259C961 00...