10.1 Operation Settings
198
If a power outage or other power loss causes an interruption in recording (while the
LED on the left side of
START
is lit), you can automatically resume recording when
the power is restored. If you are using triggers, the triggers are restored to the
[Wait-
ing for trigger]
state.
If power is lost within about three minutes after power-on, files on the removable
storage may be corrupted, and the device could be damaged. These risks can be
avoided by setting the file protection level to
[High]
.
10.1 Operation Settings
Using the Auto-Resume Function (Resume After Power Restora-
tion)
When Auto-Resume is enabled, measurement data that was in internal
buffer memory before the outage is lost when measurement resumes.
To retain measurement data from before an outage, disable Auto-
Resume (set to Off).
To retain measurement data from before an outage and resume mea-
suring automatically afterwards, enable Auto Save (p. 102).
Select from the setting options of
[Start Backup].
Setting options (
∗
. default setting)
Off
∗
Do not use the Auto-Resume Function.
On
Use the Auto-Resume Function.
File Protection Level Setting
Select from the setting options of
[File Protection]
.
Setting options (
∗
. default setting)
Low
∗
Although removable storage is recognized and
ready for recording immediately upon power-on,
if power is lost within about three minutes after
power-on, files on the removable storage may
be corrupted, and the device could be damaged.
High
The CF card or USB flash drive cannot be ac-
cessed for about 3 minutes after power-on, but
the files can be protected.
Auto-Setup is not available when
[High]
is selected (p. 158).
Summary of Contents for MR8880-20
Page 2: ......
Page 12: ...Procedure 2 ...
Page 40: ...1 3 Screen Configuration and Display 30 ...
Page 76: ...3 4 Measurement Using Setting Wizard PRESETS key 66 ...
Page 174: ...6 4 Data Management 164 ...
Page 186: ...7 7 Performing a Printer Check 176 ...
Page 196: ...8 4 Numerical Value Calculation Expressions 186 ...
Page 216: ...10 4 Making System Settings 206 ...
Page 222: ...11 3 Communication Using Commands 212 4Click Next to start install ing Installing Click ...
Page 226: ...11 3 Communication Using Commands 216 ...
Page 254: ...13 6 Setting Wizard Function PRESETS 244 ...
Page 292: ...Appendix 10 Installing Waveform Viewer A30 ...
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......