67
Select Input Channels
3.4.4
Displayed Sheet
The input channel settings can be set differently for each displayed sheet. Up to four sheets can be set.
You can set desired waveforms for displaying to different sheets and switch them.
Sheet switching (1 to 4)
• Only the following display-related settings can be set to each displayed sheet.
Analog waveform:
Display ON/OFF, waveform color, magnification, zero position, vernier, invert,
graph variable (ON/OFF, upper/lower limits)
Logic waveform:
Display ON/OFF, waveform color, display position, logic width
X-Y waveform:
Display ON/OFF, display color, X ch/Y ch waveform calculation (X ch, Y ch)
Waveform calculation: Display ON/OFF, waveform color, scale setting, upper/lower limits, graph
Common setting:
Display format
• The measurement-related settings other than above will be common to all displayed sheets.
When the range is changed, the change is applied to all displayed sheets.
•
When a settings file is saved, the settings of all displayed sheets are saved.
•
A waveform file is saved based on the setting of the sheet displayed at the time of saving. When a
waveform file is loaded, only the sheet displayed at the time of saving can be loaded because other
sheets were not saved.
3
Measurement
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......