279
FFT Analysis Modes
Cross Power Spectrum
The product of the spectra of two input signals or common frequency components of two signals
can be obtained.
Using the voltage and current waveforms as input signals, active power, reactive power and
apparent power can be obtained for each frequency.
Main uses:
To inspect common frequency components of two signals
About the Functions, refer to “13.9.2 Analysis Mode Functions” (p. 289).
Axis
Display type
Description
X axis
Linear
Displays the frequency on a linear scale
Log
Displays the frequency on a logarithm scale
Nyquist display
Displays the real-number component of the input-output ratio linearly.
Y axis
Lin-Mag
Displays the squared value of amplitude contents of analysis data linearly.
Log-Mag(logarithm)
Displays the dB value of amplitude contents of analysis data linearly. (0 dB
reference value: 1eu
2
)*
Lin-Real
Displays the squared values of the real component of analysis data linearly.
Lin-Imag
Displays the squared values of the imaginary component of analysis data
linearly.
Nyquist display
Displays the imaginary component of analysis data linearly.
* eu: engineering units that are currently set are the standard (e.g., when the unit settings is volts, 0 dB = 1 V
2
)
Waveform Example
Normal display
X axis: Log
Y axis: Lin-Mag
Normal display
X axis: Log
Y axis: Log-Mag
Normal display
X axis: Log
Y axis: Lin-Real
Normal display
X axis: Log
Y axis: Lin-Imag
Nyquist display
13
FFT
Function
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......