218
Calculation Result Judgment
10.3.1 Display Judgment Results and Signal Output
Numerical calculation judgment results are displayed on the Waveform screen.
Within the judgment threshold range: GO judgment
Out of the judgment threshold range: NG judgment (displayed in red)
When printing, judgment results for each parameter will also be printed.
When the judgment result is GO
• The GO signal is output at the GO
――
/EXT.OUT.1
―
―
―――
――
external I/O terminal.
When the judgment result is NG
• The NG signal is output at the NG
――
/EXT.OUT.2
―
―
―――
――
external I/O terminal. The NG judgment is asserted when
any channel is judged as NG.
• When the beeper is enabled, a beep sounds when a result is out of the threshold range.