172
Detailed Settings for Modules
8.10.6 Setting Model 8971 Current Unit
There is no need to change the setting since it is set when the clamp sensor is automatically
recognized.
Refer to “To Open [Each Ch] and Make a Channel Selection” (p. 164).
Mode
20A/2V
Sets this option when Model 9272-10 (20 A range), 9277, or CT6841
Clamp Sensor is connected. (Default setting)
200A/2V
Sets this option when Model 9272-10 (200 A range), 9278, CT6863, or
CT6843 Clamp Sensor is connected.
50A/2V
Sets this option when Model CT6862 Clamp Sensor is connected.
500A/2V
Sets this option when Model 9279, 9709, CT6844, CT6845, CT6846*, or
CT6865* Clamp Sensor is connected.
* When Model CT6846 or CT6865 is connected to Model 8971 Current Unit through
Model 9318 Conversion Cable, the instrument recognizes that as an 500 A AC/DC
sensor is connected; set the scaling ratio to 2.00.
Range
DC
Current measurement (
Default setting)
RMS
RMS current measurement
8.10.7 Setting Model 8972 DC/RMS Unit
Refer to “To Open [Each Ch] and Make a Channel Selection” (p. 164).
Mode
Switches between voltage measurement and RMS measurement.
DC
Voltage measurement
(Default setting)
RMS
RMS measurement
Response
The response time for RMS measurement can be set to three speeds: Fast, Normal
and Slow.
While it is normally set to
[Fast]
, this can be changed to
[Normal]
or
[Slow]
to
stabilize the measured value when the frequency is low or when severe fluctuations
are present.
Fast
Response time about 100 ms.
(Default setting)
Normal
Response time about 800 ms.
Slow
Response time about 5 s.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......