Detailed Settings for Modules
8.10.1 Settings for the Anti-Aliasing Filter (A.A.F.)
(Model 8968 High Resolution Unit)
Refer to “To Open [Each Ch] and Make a Channel Selection” (p. 164).
A.A.F
Enable the anti-aliasing filter to remove aliasing distortion. The cutoff frequency
automatically changes according to the timebase or, when the FFT function is used,
the frequency range setting.
Off
The anti-aliasing filter is disabled. (
Default setting)
On
The anti-aliasing filter is enabled.
(When external sampling is used, the antialiasing filter (AAF) is not
available.)
8.10.2 Probe Attenuation Selection
Refer to “To Open [Each Ch] and Make a Channel Selection” (p. 164).
Probe
Set this when performing measurement with a connection cable or probe.
1:1
Model L9197, L9198 or L9217 Connection Cord connected to the input
module. (
Default setting)
10:1
Model 9665 10:1 Probe connected to the input module.
100:1
Model 9666 100:1 Probe, P9000-01/-02 Differential Probe connected to the
input module.
1000:1
Model 9322 Differential Probe, P9000-01/-02 Differential Probe connected
to the input module.
Summary of Contents for MR8827
Page 19: ...14 Operation Precautions ...
Page 81: ...76 Start and Stop Measurement ...
Page 111: ...106 Manage Files ...
Page 125: ...120 Miscellaneous Printing Functions ...
Page 143: ...138 View Block Waveforms ...
Page 191: ...186 Setting Output Waveform Parameters ...
Page 291: ...286 FFT Analysis Modes Measurable Ranges With Octave Analysis 1 1 OCT 1 3 OCT ...
Page 292: ...287 FFT Analysis Modes 1 1 OCT 1 3 OCT 13 FFT Function ...
Page 293: ...288 FFT Analysis Modes 1 1 OCT 1 3 OCT ...
Page 295: ...290 FFT Analysis Modes ...
Page 309: ...304 Editor Command Details ...
Page 387: ...382 Module Specifications ...
Page 405: ...400 Dispose of the Instrument Lithium Battery Removal ...
Page 431: ...A26 FFT Definitions ...
Page 436: ......