![Hioki MR8740 Instruction Manual Download Page 367](http://html1.mh-extra.com/html/hioki/mr8740/mr8740_instruction-manual_2137324367.webp)
17.6 Input Modules Specifications
355
Chapte
r 17 S
pecifications
16
17
*: During current measurement, add the accuracy and characeristic of clamp sensor using with Model 8971.
*
2
: The instrument recognizes this as a 500 A AC/DC sensor is connected; setting the ratio of the scaling function to 2 is required.
17.6.6 8971 Current Unit
Temperature and humidity
range for guaranteed accuracy
23°C ± 5°C (73.4 ± 9.0°F), 20% RH to 80% RH (when zero adjustment is executed 30 minutes
after power on)
Product warranty period
3
year
Guaranteed accuracy period
1 year
Number of input channels
2 channels
Applicable current sensors
Model 9272-10, 9277, 9278, 9279, 9709, CT6841, CT6843, CT6844, CT6845, CT6846,
CT6862, CT6863, CT6865
Hioki current sensors with RM515EPA-10PC (Hirose) installed
(Supported conversion ratios: 2 V/20 A, 2 V/50 A, 2 V/200 A, 2 V/500 A, 2 V/1000 A*
2
)
(To connect with Model 8971, using Model 9318 Conversion Cable)
Measurement ranges
When using the Model 9272-10 (20 A), 9277, CT6841:
100 m, 200 m, 500 m, 1, 2, 5 A/div
When using the Model 9272-10 (200 A), 9278, CT6843, CT6863:
1, 2, 5, 10, 20, 50 A/div
When using the Model 9279, 9709, CT6844, CT6845, CT6846*
2
, CT6865*
2
:
2, 5, 10, 20, 50, 100 A/div
When using the Model CT6862: 200 m, 500 m, 1, 2, 5, 10 A/div
Measurement accuracy*
±0.65%f.s. (filter 5 Hz ON)
±0.85%f.s. (filter 5 Hz ON, for using with Model 9278 or 9279)
RMS accuracy*
±1%f.s. (DC, 30 to 1 kHz), ±3%f.s. (1 kHz to 10 kHz), (Sine wave input, filter 5 Hz ON)
Response time*
100 ms (rise 0
→
90%f.s.)
Crest factor
2
Temperature characteristic*
±0.075%f.s./°C
Frequency characteristic*
DC Coupling: DC to 100 kHz ±3 dB
AC Coupling: 7 Hz to 100 kHz ±3 dB (low cut-off frequency 7 Hz±50%)
Noise*
10 mAp-pmax (sensitivity range, with input shorted) (for 20 A/2V range)
Low-pass filter
OFF, 5, 50, 500, 5k, 50 k±50%(Hz)-3 dB
Input type
Unbalanced (Not insulated)
Input coupling
AC/DC/GND
Input resistance
1M
Ω
±1%
A/D resolution
12 bit
Maximum sampling rate
1 MS/s
Input terminals
Sensor connector HR10A-10R-S (HIROSE)
Operating temperature and
humidity ranges
Same as the Memory HiCorder in which the 8971 is installed
Operating environment
Same as the Memory HiCorder in which the 8971 is installed
Storage temperature and hu-
midity ranges
-10 to 50°C (14 to 122°F), 80% RH or less (no condensation)
Dimensions
Approx. 106W x 19.8H x 196.5D mm (4.17"W x 0.78"H x 7.74"D) (excluding projections)
Mass
Approx. 250 g (8.8 oz.)
Applicable Standards
Safety EN 61010
EMC
EN 61326 Class A
Accessories
For connecting the clamp sensor: Model 9318 Conversion Cable x 2, Instruction Manual
Number of the 8971 available
for Memory HiCoder
Max. 4 unit (for Model MR8740)
(cannot be used with
MR8741
)
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......