17.6 Input Modules Specifications
353
Chapte
r 17 S
pecifications
16
17
17.6.5 8970 Freq Unit
Temperature and humidity
range for guaranteed accuracy
23°C ± 5°C (73.4 ± 9.0°F), 20% RH to 80% RH
Measurement functions
Based on voltage input, measures frequency, rotation speed, power frequency, integral values,
pulse duty ratio, and pulse width.
Connection terminal
Insulated BNC terminal
Input resistance
1 M
Ω
±1%
Input capacitance
30pF±10pF
Maximum input voltage
400 VDC
Maximum rated voltage to
earth
300 VAC, DC (Measurement category II), anticipated transient overvoltage 2500 V (between each
input channel and main unit, and between input channels)
Input type
Unbalanced (input isolated from output)
Product warranty period
3
year
Guaranteed accuracy period
1 year
Frequency mode
Measurement ranges
1, 5, 10, 50, 100, 500, 1k, 5 kHz/div (f.s.=20div)
Measurement accuracy
±0.1%f.s. (except for 5 kHz/div range) ±0.7%f.s.(5 kHz/div range)
Measurement ranges
DC to 100 kHz (min. pulse width 2
μ
s)
Rotation speed mode
Measurement ranges
100, 500, 1k, 5k, 10k, 50k, 100 kr/min /div (f.s.=20div)
Measurement accuracy
±0.1%f.s. (except for 100 kr/min range) ±0.7%f.s.(100 kr/min range)
Measurement ranges
0 to 2 kr/min (min. pulse width 2
μ
s)
Power frequency mode
Measurement ranges
50 Hz (40 to 60 Hz), 60 Hz (50 to 70 Hz), 400 Hz (390 to 410 Hz) (f.s.=20div)
Measurement accuracy
±0.03 Hz (50 Hz, 60 Hz) ±0.1 Hz (400 Hz)
Integral values mode
Measurement ranges
2k, 10k ,20k ,100k ,200k ,1M counts/div
Measurement accuracy
±range/2000
Measurement ranges
DC to 100 kHz (min. pulse width 2
μ
s)
Pulse duty ratio mode
Measurement ranges
5%/div (f.s.=20div)
Measurement accuracy
±1% (10 to 10 kHz) ±4% (10k to 100 kHz)
Measurement ranges
10 to 100 kHz (min. pulse width 2
μ
s)
Pulse width mode
Measurement ranges
500
μ
, 1m, 5m, 10m, 50m, 100 ms/div (f.s.=20div)
Measurement accuracy
±0.1%f.s.
Measurement ranges
2
μ
to 2 s
Measurement resolution
Integral values mode: 2000LSB/DIV (f.s.=20 DIV)
Power frequency mode: 100LSB/DIV (f.s.=20 DIV)
Except for Integral values mode and Power frequency mode: 500LSB/DIV (f.s.=20 DIV)
Response time
Less than 40
μ
s + sampling interval of instrument in which the device is installed.
Input voltage range
±10 V, ±20 V, ±50 V, ±100 V, ±200 V, ±400 V
Threshold value
±10 V range : -10 to +10 V variable (0.1 V steps), ±20 V range: -20 to +20 V variable (0.2 V steps),
±50 V range : -50 to +50 V variable (0.5 V steps), ±100 V range: -100 to +100 V variable (1 V steps),
±200 V range: -200 to +200 V variable (2 V steps), ±400 V range: -400 to +400 V variable (5 V
steps)
Slope
Rising, falling (Frequency, Rotation speed, Power frequency, Integral values mode)
Level
High, Low (Duty ratio, Pulse width mode)
Hold
Frequency mode/ Rotation speed mode: ON/OFF (1 Hz, 0.5 Hz, 0.2 Hz, 0.1 Hz)
:When OFF is selected and the next measurement value is not determined within the waiting time
period, the frequency or the rotation speed will be determined using the value which is calculated
based on the time period between the timing of the last measurement and the timing of the sampling.
Put it to 0 when the measurement value is below the fixed value.
Smoothing
OFF, ON* (Frequency mode,Rotation speed mode)
*:The permissible smoothing frequency is up to 10 kHz.
Low-pass filter
OFF, 5, 50, 500, 5k, 50 kHz
Input Coupling
DC, AC (Low frequency cut-off in AC-coupled mode : 7 Hz)
Frequency dividing function
Setting range: 1 to 4096, by one step (Frequency ,Rotation speed ,Integral values mode)
Integration start timing
Start, Trigger (Integral values mode)
Integration limit
Hold, Loop (Integral values mode)
Operating temperature and
humidity ranges
Same as the Memory HiCorder in which the 8970 is installed
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......