![Hioki MR8740 Instruction Manual Download Page 363](http://html1.mh-extra.com/html/hioki/mr8740/mr8740_instruction-manual_2137324363.webp)
17.6 Input Modules Specifications
351
Chapte
r 17 S
pecifications
16
17
17.6.3 8968 High Resolution Unit
Temperature and humidity
range for guaranteed accuracy
23 ± 5°C (73.4 ± 9.0°F), 20 to 80% RH (when zero adjustment is executed 30 minutes after power
on)
Product warranty period
3
year
Period of guaranteed accuracy
1 year
No. of input channels
2 channels
Measurement ranges
5, 10, 20, 50, 100, 200, 500 mV, 1, 2, 5, 10, 20 V/div
Measurement accuracy
±0.3% f.s. (Filter 5 Hz On, after zero adjustment)
Temperature characteristic
±0.045% f.s./°C
Frequency characteristic
DC coupling: DC to 100 kHHz -3 dB
AC coupling: 7 Hz to 100 kHz -3 dB (low-cut frequency: 7 Hz±50%)
Noise
500
μ
Vp-p (typ), 1 mVp-p (max) (sensitivity range, with input shorted)
Common mode rejection ratio
80 dB minimum (at 50/60 Hz and with signal source resistance 100
Ω
maximum)
Low-pass filter
Off, 5±50%, 50±50%, 500±50%, 5 k±50%, 50k±50%(Hz)-3 dB
Anti-aliasing filter
Cutoff frequency (fc) 20, 40, 80, 200, 400, 800, 2k, 4k, 8k, 20k, 40k (Hz)
(Automatic setting when the anti-aliasing filter is ON)
Attenuation property at 1.5 fc, -66 dB or higher
Input type
Unbalanced (floating)
Input coupling
AC/DC/GND
Input resistance
1 M
Ω
±1%
Input capacitance
30 pF±10 pF (at 100 kHz)
A/D resolution
16 bit
Maximum sampling rate
1 MS/s
Input terminals
Insulated BNC terminal
Maximum input voltage
400 V DC
Maximum rated voltage to
earth
300 V AC/DC (between input channels and chassis, and between input channels)
Measurement category II (anticipated transient overvoltage 2500 V)
Operating temperature and
humidity
Same as the host Memory HiCorder
Operating environment
Same as the host Memory HiCorder
Storage temperature and hu-
midity
Temperature -10°C to 50°C (14°F to 122°F) Humidity 80% RH or less (non-condensating)
Dimensions
Approx. 106 mmW × 19.8 mmH × 207.5 mmD (4.17"W × 0.78"H × 8.17"D)
Mass
Approx. 250 g (8.8 oz)
Effect of radiated radio-fre-
quency electromagnetic field
±15% f.s. (max) at 3 V/m
Effect of conducted radio- fre-
quency electromagnetic field
±20% f.s. (max) at 3 V (at 100 mV/div range, 1 VDC input)
Applicable standards
Safety
EN61010
EMC
EN61326 Class A
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...