7.10
Register the Waveform in the U8793 Arbitrary
Waveform Generator Unit
179
6
Chapte
r 7 Utility Functions
7
You can register waveform in the Model U8793. Registered waveforms can be output from the Model
U8793.
7.10 Register the Waveform in the U8793
Arbitrary
Waveform Generator Unit
Procedure
To open the screen: press the
CHAN
key to open the Channel screen, and then select
[Generator]
sheet
1
Set the type to arbitrary waveform.
Move the cursor to
[Type
]
on the settings screen, and select
[Arbitrary]
using the
F
keys.
2
Register the waveform.
Move the cursor to the item listed below
[Arbitrary]
on the settings screen.
3
Select a reading source of the waveform.
Select
Selections
Description
Resister:
From File
Registers from data saved in the media.
Register:
From Meas.
Registers the data measured using memory function. You can also
register after reading the MEM files from media to the instrument.
↑
or
↓
Select arbitrary waveform data registered in the Model U8793.
(Use when selecting/deleting output waveforms.)
Delete
Deletes the data registered in the U8793 memory.
If eight waveforms are already registered in the memory, delete
any of them and then register.
Summary of Contents for MR8740
Page 2: ......
Page 12: ...Contents x ...
Page 26: ...Operating Precautions 14 ...
Page 128: ...5 5 Miscellaneous Printing Functions 116 ...
Page 196: ...7 12 Setting Output Waveform Parameters 184 ...
Page 234: ...9 5 Numerical Calculation Type and Description 222 ...
Page 248: ...10 3 Waveform Calculation Operators and Results 236 ...
Page 302: ...12 8 FFT Analysis Modes 290 ...
Page 316: ...13 6 Editor Command Details 304 ...
Page 416: ...Appendix 4 FFT Definitions A26 ...
Page 421: ......
Page 422: ......
Page 423: ......
Page 424: ......