![Hioki MR8740-50 Instruction Manual Download Page 219](http://html1.mh-extra.com/html/hioki/mr8740-50/mr8740-50_instruction-manual_2137326219.webp)
213
External Input and Output
Trigger output (TRIG.OUT)
The instrument outputs the signal when it is triggered. You can use this signal to control multiple
instruments, achieving synchronous operation.
How to output the signal
1
Connect each of the TRIG OUT and GND terminals to an external signal-inputting device
with wires.
Refer to “2.3 Connecting the External Control Terminals” in Quick Start Manual.
2
> [System] > [External terminal]
3
3
In the
[External trigger]
area, click the
[Trigger out]
box, and then from the list, choose a
signal output method.
Pulse
Outputs a low-level signal, and then switches it to the high level after the specified time
has elapsed.
Level at time
Continuously outputs a low-level signal after triggered during the measurement.
When triggered, the instrument outputs a pulse wave, which switches from the high level (4.0 V to 5.0 V) to the
low level (0 V to 0.5 V).
Output signal
Open-drain output (with voltage output), active low*
Output voltage range
High level:
4.0 V to 5.0 V
Low level:
0 V to 0.5 V
Pulse width
With the pulse
setting:
2 ms ±1 ms
When the level
setting:
(Sampling rate) × (Number of data points after trigger) or longer
Maximum input
voltage
50 V DC, 50 mA, 200 mW
*: The instrument is triggered when the signal voltage level switches from the high level to the low level.
12
Externally Controlling the Instrument
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......