107
Triggering the Instrument Using Analog Signals (Analog Trigger)
1.
[Level]
trigger
When an input signal crosses the specified level in the positive or negative direction, an analog
trigger is generated.
Level
Input waveform
Slope:
[ ]
[ ]
Setting
Description
Level
–f.s.
to
+f.s.
Default:
0
Allows you to enter a level of the level trigger.
Slope
With
OR
The level-trigger condition is satisfied when a waveform crosses
the threshold value (level) in the positive direction.
The level-trigger condition is satisfied when a waveform crosses
the threshold value (level) in the negative direction.
With
AND
HIGH
The level-trigger condition is satisfied when a waveform is higher
than the threshold value (level).
LOW
The level-trigger condition is satisfied when a waveform is lower
than the threshold value (level).
Event
With
OR
1
to
4,000
Allows you to enter the number of events.
The instrument counts the number of times the level-trigger
condition is satisfied. An analog trigger is generated when the
count reaches the specified number of events.
With
AND
Not available
Filter
Off
,
10
to
10,000
Allows you to enter a filter width in the number of samples. Only
after the level-trigger condition is continuously satisfied during
the specified period, an analog trigger is generated. Specifying
this option prevents the instrument from being unintentionally
triggered due to noise.
With the
[Filter
]
setting
Specifying the filter duration prevents the analog trigger from being unintentionally generated due
to noise, allowing it to be generated only after the level-trigger condition continues to be satisfied
during the specified duration (period) or longer.
Filter
Level at time
Noise does not generate any analog trigger.
5
Configuring the
Trigger Settings
Summary of Contents for MR8740-50
Page 2: ......
Page 6: ...Contents iv...
Page 40: ...34 Enlarging a Part of the Waveform Zoom Function...
Page 86: ...80 Configuring Generator Module Specific Settings...
Page 128: ...122 Automatically Specifying the Trigger Level Auto trigger Level...
Page 164: ...158 Numerical Calculation Types and Descriptions...
Page 182: ...176 Displaying Waveforms...
Page 186: ...180...
Page 214: ...208 Operating the Instrument With a Browser Installed in a Computer...
Page 224: ...218 External Sampling EXT SMPL...
Page 232: ...226 Information for Reference Purposes...
Page 236: ...230 Index...
Page 237: ......
Page 238: ......