77
Configuring Module-Specific Settings
5
Tap the
[Calibration]
box, and then choose a calibration setting from the list.
Enabling this setting automatically calibrates the module or synchronizes the channels at the start of
measurement. The synchronization between channels allows integration calculations to synchronize with each
other.
Off
Does not calibrate the module nor synchronize the channels.
On
Calibrates the instrument and synchronizes the channels.
Synchronization
Only synchronizes the channels.
•
It takes about 150 ms to calibrate the module. The instrument cannot measure any input signals during
calibration.
• If the channels are synchronized, the instrument sends a signal that interrupts the integration to each
module at the start of measurement; thus, the instrument has to wait until the first integration finishes.
The wait time required for this process stands at the sum of 10 ms and integration time*.
*: The integration time varies depending on the NPLC setting.
Even when synchronization is not performed, the wait time described above is required for measurement
performed immediately after the settings of Model MR8990 Digital Voltmeter Unit has been changed. No
wait time is required when the measurement is performed without any setting changes
• When
[Calibration]
is set to
[Off]
(default setting), execute calibration manually.
See “2.12 Executing Calibration (For the Instrument With Model MR8990 Installed)” in the Quick Start
Manual.
6
Tap
[Close]
.
The setting dialog box closes.
Advanced Functions
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...