58
Converting Input Values (Scaling Function)
When using
[dB]
: Specify a physical quantity per input signal (ratio) in decibels.
Tap the
[Input dB]
box, and then enter an inputted signal value in decibels.
Tap the
[Output dB]
box, and then enter an scaled value in decibels.
−200
to
+200
You can specify values to five or less significant figures.
Setting example:
Converting (scaling) an input value of 40 dB into 60 dB
Input dB
:
40
Output dB
:
60
The instrument specifies a conversion ratio corresponding to inputted values in decibels. The offset
value is set at zero.
When using
[Rating]
(for Model U8969 Strain Unit only)
Tap the
[Capacity]
box, and then enter the rated capacity of a strain gauge converter to be used.
Tap the
[Output]
box, and then enter a rated output of a strain gauge converter to be used.
+1.0000E-9
to
+9.9999E+9
You can specify values to five or less significant figures.
Specify the parameters such that the quotient of the rated capacity divided by two times the rated
output is less than or equal to 9.9999E+9.
For the rated capacity and rated output, see an inspection record of a strain gauge converter to be
used.
Setting example: To display results measured with a strain gauge converter that has a rated capacity of
20 G and rated output of 1000 μV/V as figures in gravities (G)
Unit: G
Rated capacity: 20
Rated output: 1000
The upper and lower display values of a waveform also automatically change according to the
changes made in the scaling settings.
4
Tap the
[Units]
box, and then enter a unit used for scaled values.
(Number of characters: up to 7)
To copy the scaling setting to another channel
See “3.6 Copying Settings (Copy Function)” (p. 64).
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...