56
Converting Input Values (Scaling Function)
3
Tap the
[Method]
box, and then choose a setting method from the list.
Ratio
Allows you to specify a conversion ratio and an offset.
2-Point
Allows you to specify two scaling-reference points.
Sensor
Allows you to choose a model name and measurement range of a connected current
sensor or differential probe.
Output rate
Allows you to choose an output rate (ratio) of a current sensor or a division ratio of a
voltage dividing probe.
dB
Allows you to specify an input value in decibels and value after scaled.
Rating
Allows you to specify a rated capacity and rated output according to an inspection
record of a strain gauge converter to be used. (For Model U8969 Strain Unit only)
When using
[Ratio]
Enter a conversion ratio in the
[Ratio]
box.
−9.9999E+9
to
9.9999E+9
Enter an offset value in the
[Offset]
box.
−9.9999E+19
to
9.9999E+19
To convert inputted voltage values into current values
Specify a physical quantity per one volt of an inputted
signal (conversion ratio: eu/V), an offset value, and a
measurement unit to be used. The instrument converts
(scales) measured values acquired in volts into that in the
specified measurement unit. (eu: engineering unit)
Example:
Ratio: Change in amperes per volt; offset value: B
Unit: A
(A)
B
Scaling values using a slope
(conversion ratio) and offset value
(V)
When using
[2-Point]
Enter a value in each of the following fields:
[Input1]
,
[Input2]
,
[Scale1]
,
[Scale2]
.
Enter two voltage values of an inputted signal in the
[Input1]
and
[Input2]
boxes.
Enter each of two scaled value in the
[Scale1]
and
[Scale2]
boxes.
−9.9999E+29
to
9.9999E+29
The instrument converts (scales) measured values in
volts into values in the specified measurement unit.
Example:
Input1
,
Input2
Scale1
,
Scale2
V
H
: Higher potential
point
A
H
: Value for the higher
potential point
V
L
: Lower potential
point
A
L
: Value for the lower potential
point
Unit: A
(A)
A
H
A
L
V
L
V
H
Scaling values using a conversion ratio
and offset value, both of which are
calculated using two points
Values after scaled
Actual measured values
(V)
When changing the value with the
[Ratio]
setting, the instrument does not change the values for V
L
or
V
H
, but the values for A
L
and A
H
, all of which have been specified with
[Method]
set to
[2-Point]
.
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...