308
Saving Data
12.8 Saving Data
You can save all CAN frame data measured.
File format
CAN frame data is saved as files different from waveform files. Data can be saved in binary form or
text form.
Type
Form
Extension
Description
CAN binary
Binary
CLG
Saves a waveform file and an index file as a
package (Saving either of them is unavailable).
Loading an index file enables the instrument to load
waveforms and frame data into the internal memory.
CAN text
Text
CSV, TXT
Only manually saving is available.
CAN binary file
(1) Saving by pressing the
SAVE
key (manual save)
When a CAN binary file is saved, a CAN binary file can be simultaneously saved.
Tap the
[CAN]
button to specify the file to be saved.
Off
Saves the waveform file only.
On
When the CAN frame data exists, the CAN log file and index file, as well as the
waveform file, are saved.
(2) Automatic saving, real-time saving
Automatically saves the waveform binary file. For the real-time saving, automatically saves the
CAN binary file.
CAN text file
CAN text files can be saved by pressing the
SAVE
key (manual save) only.
Tap the
[Type]
button, and choose
[CAN(Text)]
.
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...