
263
Configuring Dual Sampling Settings
11.2
Configuring Dual Sampling Settings
> [Status] > [Condition]
1
5
2
3
4
1
Tap the
[Dual Sampling]
box to set to
[On]
.
The following settings change automatically.
Measurement
method
Envelope
Realtime save
Off
Memory division
On
Overlay
Off
2
Tap the
[Mode]
box, and then from the list, choose a recording mode (behavior after
finishing trend waveforms).
Single
Finishes recording after the set recording length of trend waveforms has been
recorded.
Repeat
Clears waveforms after the recording length of trend waveforms have been recorded
and records waveforms once again from the beginning. At this time, instantaneous
waveforms are also recorded from the start block. You can stop recording by pressing
the
STOP
key
3
Set the sampling speed and the number of points for the trend waveforms.
The instrument measures trend waveforms using the envelope method (records the maximum and minimum
values per sample).
At this time, it performs oversampling of the input signals at the sampling speed specified for the instantaneous
waveform recording.
See “Using the envelope” (p. 9).
Dual Sampling
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...