138
Triggering the Instrument Using Logic Signals (Logic Trigger)
5.7 Triggering the Instrument Using Logic Signals
(
Logic Trigger)
The section explains the logic trigger function.
•
Input signals acquired across the logic channels serve as a trigger source.
• You can set a trigger pattern and trigger satisfaction condition by choosing between logical AND
and OR operations. When the logic-trigger conditions are satisfied, a logic trigger is generated.
•
With the trigger filter setting, no logic triggers are generated until the logic-trigger condition is
continuously satisfied during the specified filter.
> [Trigger] > [Source]
1
3
2
1
Tap the
[Condition satisfied]
box, and then choose a logic-trigger satisfaction condition from
the list.
Off
Disables the logic trigger.
OR
The logic-trigger condition is satisfied when any one of logic input signals matches the
trigger pattern.
AND
The logic-trigger condition is satisfied when all logic input signals match the trigger
pattern.
2
Tap the
[
Filter]
box, and then choose a sampling count of the filter from the list.
Only after the level-trigger condition is continuously satisfied during the specified period, an analog trigger is
generated.
Configuring the trigger filter setting prevents a logic trigger from being unintentionally generated due to noise.
Off
Disables the trigger filter.
10
to
10000
Enables the trigger filter.
Allows you to enter a filter width in the number of samples.
When the envelope is used
Off
Disables the trigger filter.
1 ms
,
10 ms
Enables the trigger filter.
HIOKI MR6000A966-03
Summary of Contents for Memory HiCorder MR6000
Page 2: ...HIOKI MR6000A966 03...
Page 56: ...50 Displaying Measured Values HIOKI MR6000A966 03...
Page 152: ...146 Triggering the Instrument Using CAN Signals CAN Trigger HIOKI MR6000A966 03...
Page 194: ...188 Numerical Calculation Types and Descriptions HIOKI MR6000A966 03...
Page 266: ...260 Configuring the Display Settings HIOKI MR6000A966 03...
Page 391: ...HIOKI MR6000A966 03...
Page 392: ...HIOKI MR6000A966 03...