95
Activating a Trigger at a Set Interval
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
2.6 Activating a Trigger at a Set Interval
Interval triggers
This section describes how to activate a trigger at a set interval.
You need to set repeat recording set to
[ON]
beforehand.
See “1.2 Setting Measurement Conditions” (p. 10).
>
>
1
1
2
2
1
Under
[Interval trigger]
, select the interval trigger conditions.
If you wish to give priority to the interval trigger, use the
[OR]
If you wish to give priority to other triggers, use the
[AND]
OFF
Does not use the interval trigger.
OR
Uses the interval trigger based on OR logic.
AND
Uses the interval trigger based on AND logic.
2
Set the interval for the interval trigger.
DD, HH:MM:SS
The trigger will simultaneously activate when measurement starts. Subsequently, the trigger will activate at
every interval trigger interval.
2
Trigger Function
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...