28
Configuring Input Channels
5
Under
[Filter]
, select the cutoff frequency.
Auto
,
120 Hz
,
60 Hz
,
30 Hz
,
15 Hz
,
8 Hz
,
4 Hz
When
[Auto]
is selected, the low-pass filter’s cutoff frequency will automatically be set as described
in the following table based on the set data refresh interval:
Data refresh interval
Cutoff frequency
Data refresh interval
Cutoff frequency
1 ms
120 Hz
200 ms
4 Hz
2 ms
60 Hz
500 ms
4 Hz
5 ms
30 Hz
1 s
4 Hz
10 ms
15 Hz
2 s
4 Hz
20 ms
8 Hz
5 s
4 Hz
50 ms
4 Hz
10 s
4 Hz
100 ms
4 Hz
6
On the
[Wave+Set]
screen, press the
ENTER
key while
[Auto-bal]
is selected on the bottom
right of the screen.
Auto-balancing will be performed for all channels of the Strain Unit.
Perform this step under the following conditions:
• Turn on the instrument and wait 30 minutes.
• Connect a strain gage or strain gage-type converter to the module; however, do not apply any
loads, including vibration.
• Connect the B and D inputs terminals with one another when measuring voltage.
Auto-balancing cannot be performed while measurement is in progress.
Key operations are ignored while auto-balancing is in progress.
Auto-balancing can also be performed using the
[Auto-balance]
button on the top right of the channel list screen. Press the
ENTER
key to display the settings window.
Select the checkboxes for the channels for which you wish to
perform auto-balancing. Then press the
ENTER
key while
[Execute]
is selected.
Press the
ENTER
key while
[All]
is selected to select or cancel
auto-balancing for all channels at once.
See “1.7 Configuring Channels in a List” (p. 50).
Perform auto-balancing again in the following circumstances:
• When the input types have changed
• When the range has changed
• When module connections have changed
• When the strain gage or strain gage-type converter has changed
• When the instrument’s power has been cycled
• When the instrument’s settings have been initialized
• When the ambient temperature has changed abruptly (when the zero position may have drifted)
If auto-balancing fails, check the following:
• Is the strain gage or strain gage-type converter in a no-load state?
• Is the strain gage or strain gage-type converter connected properly?
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...