21
Configuring Input Channels
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
Measuring temperature (with thermocouples)
This section describes how to configure settings on the individual settings screen when measuring
temperature using thermocouples.
You can use
[Input]
on the settings list screen to configure the settings. (See p. 50.)
Applicable modules: U8550, U8551, U8552, LR8530, LR8531, LR8532
>
>
3
3
4
4
1
1
5
5
6
6
7
7
2
2
1
Select the module (Unit) and channel to configure and select the checkbox.
Measurement will not be performed for channels whose checkboxes are not selected.
2
Select the waveform display color.
×
(OFF), 24 colors
Select
[×]
if you wish to measure the channel but not to display its waveform or numerical values on the
screen.
3
Set the input type to
[Tc]
.
4
Under
[Range]
, select the measurement range as appropriate for the temperature being
measured.
100°C
,
500°C
,
2000°C
B thermocouples cannot be selected for the 100°C and 500C° ranges.
When using B thermocouples, set the range to 2000°C first.
5
Under
[Type]
, select the type of thermocouple you’re using.
K
,
J
,
E
,
T
,
N
,
R
,
S
,
B
*,
C
*:
[B]
can be selected when using the 2000°C range.
See “Measurable temperature range” (p. 22).
1
Settings and Operation
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...