147
Performing Numerical Calculations
Unit
Measure
Channel
Trigger
Alarm
Calculation
System
Configuring numerical calculations
>
>
2
2
3
3
4
4
1
1
5
5
6
6
7
7
1
Under
[Numerical calculation]
, set the numerical calculation function to
[ON]
.
OFF
,
ON
2
Under
[Time split calculation]
, select the file segmentation method.
Disable
Saves the last numerical calculation results.
Enable
Allows you to set the interval at which to perform the calculation under
[Split time]
.
The calculation will be performed after a start trigger, and the results saved, at the set
interval.
For example, if the segment time is set to 10 min., the calculation will be performed, and the
calculation results saved, every 10 min.
(When
[Time split calculation]
is set to
[Enable]
)
Under
[Split time]
, set the time interval at which to perform the calculation.
0
Days
00:01
to
30
Days
23:59
If
[Time split calculation]
is set to
[Enable]
, calculation results will be saved for each time interval.
Segment time
Start
measurement
Stop
measurement
Under
[Split time]
, set the time interval at which to
perform the calculation.
Calculation results will be saved for the specified amount
of time.
The value 0 d 0 hr. 0 min. cannot be specified as the
segment time. The value will be automatically changed to
0 d 0 hr. 1 min.
If the format for numerical calculation results in the auto save settings has been set to
[OFF]
, the
[Time split calculation]
setting will not be available.
See “Auto save (real-time save)” (p. 106).
Set the format for numerical calculation results to
[Text]
.
6
Numerical and W
aveform Calculations
Summary of Contents for LR8450
Page 68: ...64 Observing Waveforms Alarm...
Page 72: ...68 Observing Waveforms 3 Numerical calculation Numerical calculation results...
Page 131: ...128 Acquiring Data with a Computer PC...
Page 175: ...172 Controlling the System...
Page 233: ...230 Controlling the Instrument with Communication Commands...
Page 271: ...268 Plug in Module Specifications...