3.4 Setting Configuration 2 – Input Channel Settings
58
Configure these settings for each voltage measurement channel.
Voltage Measurement Settings
1
Select a channel (
CH 1
to
CH 10
) for the
[Channel]
setting contents.
2
To enable measurement on the channel,
select On to display the check mark (
).
3
Select
[Voltage]
for the
[Input]
(input type)
setting.
4
Select the appropriate measurement ampli-
tude range for the
[Range]
setting.
Setting options:
10 mV*, 100 mV, 1 V, 10 V, 20 V, 100 V, 1-5 V
(
∗
default setting)
1
2
3
4
: On
: Off
Instrumentation and Control Device Measurements
For instrumentation control signal measurements, select
[1-5 V]
. This range corre-
sponds to the 4-20 mA output standard for instrumentation control. This setting automat-
ically sets 1- and 5-volt lower and upper limits, respectively, in the 10 V range.
Measurement accuracy is ±10 mV ±5 mV. To measure 4-20 mA current, connect a 250
Ω
shunt resistor b and – analog input terminals (p. 36).
Use the scaling function to scale measurements to any arbitrary unit value for display (p.
69).
If the Allowable Measurement Range of each Range is Significantly
Exceeded
“OVER” or “-OVER” is displayed for the numerical value or cursor value and the calcu-
lated value or saved data is handled as a full scale value.
Summary of Contents for LR8432-20
Page 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Page 2: ......
Page 22: ...Operating Precautions 14...
Page 50: ...2 5 Zero Adjustment 42...
Page 88: ...3 7 Viewing All Channel Settings in a List 80...
Page 104: ...4 5 About Recording Operation 96...
Page 154: ...7 4 Waveform Calculation Formula 146...
Page 164: ...8 3 Making System Settings 156...
Page 232: ...Appendix 12 Heat Flow A44...
Page 236: ...Index Index 4...
Page 237: ......
Page 238: ......