3.3 Setting Configuration 1 – Recording Settings
44
Specify how you want recording to start and
stop. The following methods are available.
• Record continuously:
Record continuously between starting and stop-
ping measurement.
• Specified length (time span):
Record only for a specified time span (can be set
to automatically repeat) (p. 45)
Selecting the Recording Start/stop Method
Continuous
Recording
Repeat Re-
cording (speci-
fied lengths)
Start
Measurement
Stop
Measurement
To record continuously,
Enable continuous recording (set
[Cont]
to
[On]
). (default setting)
Recording continues until you press the
START/STOP
key, or until a “stop trigger” event occurs.
When auto save is enabled for measurement data,
the saving method depends on whether [Deleting] or
[Split Save] (segmented save) is selected (p. 46).
To record for a specified length (time span),
1
Set
[Cont]
to
[Off]
.
2
Specify the duration of one recording length
(time span).
Setting options:
d (days)
0
to
499
, h (hours)
0
to
23
,
min (minutes)
0
to
59
, s (seconds)
0
to
59
1
2
Recording proceeds for the specified length. Record-
ing stops early if you press the
START/STOP
key be-
fore the recording length has elapsed, or if a “stop
trigger” event occurs.
Change 1’s digit
Change 10’s digit
Maximum recording length
and data samples
Indicates the maximum record-
ing time to internal memory, and
the maximum number of data
samples. These values are limit-
ed by the specified recording
length and the number of mea-
surement channels used.
See:
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...
Page 203: ......