1.5 Basic Operation
24
You can verify the input status and display range while making settings on the
Setting screen. This is not available while measuring.
Verifying the Input Level (Level Monitor)
Current Input
Signal Level
and Channel
Number
Level Monitor
Display Channel and Gauge
Shows the gauge for the current-
ly selected channel. The gauge
display is linked to the measure-
ment and display range settings.
Press the
CH
/
keys to dis-
play the gauge for another chan-
nel.
Changing the Monitor Display
Select whether to display all channels, or only se-
lected channels.
[All Ch]/[SlctCh]/[Value]
•
[All Ch]
The channels displayed in the level monitor are
those with a check mark (
) displayed next to
the channel number in the settings.
•
[SlctCh]
(Selected Channels)
The level monitor displays the channel selected
on this screen.
•
[Value]
The input values of all channels are displayed
numerically.
Displayable Waveform Range
This is determined by the waveform display range
setting.
Input Waveform Amplitude
Indicates real-time input ampli-
tude.
Using the Scaling function
(p. 58), measurement units
can be converted as desired
for display.
Summary of Contents for LR8431
Page 2: ......
Page 20: ...Operating Precautions 12...
Page 44: ...2 5 Zero Adjustment 36...
Page 74: ...3 7 Viewing All Channel Settings in a List 66...
Page 88: ...4 5 About Recording Operation 80...
Page 120: ...6 6 Transferring Data to a Computer USB Drive Mode 112...
Page 126: ...7 2 Numerical Value Calculation Expressions 118...
Page 136: ...8 3 Making System Settings 128...
Page 144: ...9 5 Synchronous Measurements with Multiple Instruments 136...
Page 154: ...Chapter 10 Specifications 146...
Page 198: ...Appendix 11 Pulse Input Circuit Diagram A40...
Page 202: ...Index Index 4...
Page 203: ......