7.1 Operation Settings
146
If a power outage or other power loss causes an interruption in recording (while the LED
on the left side of
START
is lit), you can automatically resume recording when the power
is restored. If you are using triggers, the triggers are restored to the
[Waiting for trigger]
state.
Start Backup
(Auto-
Resume)
Key Setting Procedure
Select the
[System]
Screen.
Move to a setting item.
Apply
Open the setting options
for the item to be set.
Select from the listed op-
tions.
1
2
7.1
Operation Settings
Using the Auto-Resume Function (Resume After Power Restora-
tion)
When Auto-Resume is enabled, measurement data that was in internal
memory before the outage is lost when measurement resumes. To retain
measurement data from before an outage, disable Auto-Resume (set to
Off).
To retain measurement data from before an outage and resume measuring
automatically afterwards, enable Auto Save (p. 127).
Setting options:(
∗
: default setting)
Off
∗
Do not use the Auto-Resume Function.
On
Use the Auto-Resume Function.
Summary of Contents for LR8400-20
Page 1: ......
Page 2: ......
Page 4: ......
Page 26: ...Operating Precautions 14...
Page 96: ...3 9 Viewing and Editing with the All Channel Settings List 84...
Page 112: ...4 3 Marking Waveforms and Searching Marks Search Function 100...
Page 156: ...6 5 Transferring Data to a PC USB Drive Mode 144...
Page 238: ...10 8 About Communications Commands 226...
Page 256: ...11 3 LR8501 Universal Unit Specifications 244...
Page 262: ...12 3 Disposing of the HiLogger 250...
Page 296: ...Index Index 4...
Page 297: ......
Page 298: ......
Page 299: ......
Page 300: ......