9.4 Error Displays
95
3
9
Chapter 9 M
aintenanc
e and Service
The display appears as follows when an error occurs on the logger.
9.4 Error Displays
Logger Error Displays
Error Displays
Meaning
Remedies and References
Calibration data error:
A fault occurred with the internal
calibration data.
Inspection and repair is required.
Please contact your dealer or Hioki
representative.
See:
Microcomputer error:
A fault occurred in microcomputer
ROM/RAM.
Data recording error:
A fault occurred in recording data
or accessing settings.
Battery voltage is too low for nor-
mal logger operation.
Replace the battery.
See:
"2.1 Installing (or Replacing)
the Battery" ( p.17)
or
A measured value is out of range.
Out-of-range values cannot be dis-
played.
[OF]
or
[UF]
is displayed when this
data is imported by the Utility Pro-
gram.
• The sensor plug is inserted
incorrectly.
• The sensor plug is not inserted
all the way in.
• The sensor is damaged.
• The logger is damaged.
Verify the correct plug orientation,
and insert it as far as possible.
If the values are not displayed despite
these measures, the sensor and log-
ger need to be inspected and re-
paired.
Please contact your dealer or Hioki
representative.
See:
[ERROR]
is displayed when this
(faulty) data is imported by the Utility
Program.
LR5000 Utility Program Error Displays
Error Displays
Meaning
Remedies and References
OF
A measured value is out of range.
Out-of-range values cannot be dis-
played.
UF
HIOKI LR5011B980-04
Summary of Contents for LR5011
Page 2: ...HIOKI LR5011B980 04...
Page 6: ...Contents iv HIOKI LR5011B980 04...
Page 66: ...4 8 Printing Recorded Data 60 HIOKI LR5011B980 04...
Page 82: ...6 4 Extracting Data 76 HIOKI LR5011B980 04...
Page 102: ...9 4 Error Displays 96 HIOKI LR5011B980 04...
Page 108: ...Index Index 4 HIOKI LR5011B980 04...
Page 109: ...HIOKI LR5011B980 04...
Page 110: ...HIOKI LR5011B980 04...
Page 111: ...HIOKI LR5011B980 04...
Page 112: ...HIOKI LR5011B980 04...