3.2 Measuring Insulation Resistance
62
3.2 Measuring Insulation
Resistance
Observe the following to avoid electric
shock and short circuits.
A. Do not use the
instrument if the
shutter is broken.
B. Check Table 1
before connecting test leads to the
instrument.
C. Check to see if the object under test is
not live or electrically charged using a
high-voltage detector or other similar
instrument, before connecting test
leads to it.
Table 1
Check item
Result
Action
Are the
mark
and
key lamp
off?
Off
Connect test leads to the in-
strument and check C. above.
If safe to proceed, connect the
test leads to the object under
test.
Go to Table 2.
Blinking
Press the
key to stop
voltage generation.
Table 2
Check item
Result
Action
Are the
mark
and
key lamp
blinking?
Not blinking
Measurement may be com-
menced
Blinking
Immediately disconnect the
test leads from the object un-
der test and turn off power to
the object or discharge the
electric charge using a dis-
charge rod.
Shutter