Overview
Overview
The device is the test fixture to be used by connecting to the impedance analyzer of Hioki.
It performs two-terminal measurement of lead components and SMDs.
Measurable DUT size
Lead components
Axial
A
B
C
A
Lead interval
(Component body length)
1 mm to 25 mm
B
Lead length
2 mm to 10 mm
C
Height to lead*
1
2.5 mm or shorter
Lead wire diameter
0.5 mm to 2 mm
Radial
D
E
D
Lead interval
2 mm to 26 mm
E
Lead length*
1
2 mm or longer
Lead wire diameter
0.5 mm to 2 mm
SMD*
2
H
F
G
F
Component length
1.6 mm to 23 mm
G
Component width
0.8 mm or longer
H
Component height
0.65 mm or longer
Supported SMD size: EIA0603 (JIS1608) or larger
*1: The component’s H-side lead must not make contact with the fixture’s L-side electrode.
*2: The electrode on the bottom surface of the H side of the component must not make contact with the fixture’s
L-side electrode.