![Hioki IM7580 Instruction Manual Download Page 98](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306098.webp)
90
Set Measurement Conditions for Sweep Points
3
4
The numeric keypad can be
used for input.
3
Set the voltage or current with
/
or the numeric
pad.
(With the numeric pad, press
[dBm]
.)
Measurement signal
mode
Settable range
Power (P) mode
-40.0 dBm to +7 dBm
(resolution, 0.1 dBm)
Voltage (V) mode
4 mV to 1001 mV
Current (I) mode
0.09 mA to 20.02 mA
[C]
Repeats the input.
4
Press
[EXIT]
to close the measurement signal level
setting screen.
5
Press
[EXIT]
to close the advanced settings
screen.
When a measured value is outside the guaranteed accuracy range,
REF VAL
is displayed in the
error display area.
In this case, the measurement signal level is considered to be low. Check the guaranteed
accuracy range and change the measurement conditions or consider the measured values as
values for reference.
See “Measurement range” (p. 274).
4.4.3
Setting the Measurement Speed
Change the measurement time.
When the measurement speed is set to
[SLOW]
or
[SLOW2]
, the measurement accuracy
improves.
•
Perform calibration or compensation again if there is a change in the measurement speed.
See “5 Calibration and Compensation” (p. 139).
•
Measurement time varies with the measurement conditions.
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......