![Hioki IM7580 Instruction Manual Download Page 96](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306096.webp)
88
Set Measurement Conditions for Sweep Points
4.4 Set Measurement Conditions for Sweep Points
Sets the measurement conditions for sweep points.
Setting is possible from the setting and editing of sweep points.
Reference: “4.3.3 Normal Sweep” (p. 82), “4.3.4 Segment Sweep, Segment Interval Sweep”
4.4.1
Setting Measurement Signal Frequency
Sets the measurement signal frequency.
1
3
Input method can be switched
between
[DIGIT]
and
[10-KEY]
.
/
When setting with (each digit)
1
Set the frequency with
/
or with the numeric
keypad.
Settable range 1 MHz to 300 MHz
/
When setting with (each digit)
Holding down a digit key changes the value continuously.
Settable range 1 MHz to 300 MHz
[×10]
Sets the measurement frequency to 10×.
[/10]
Sets the measurement frequency to
1/10×.
1
2
To set the frequency with the
numeric keypad
Changing the unit:
M
(mega)/
k
(kilo)
To set the frequency with the numeric keypad
[C]
Repeats the input.
•
Instrument key is enabled if a numerical value is input.
•
Frequency is set on pressing the instrument key.
•
If the value is set to more than 300 MHz:
The value is automatically set to 300 MHz.
•
If the value is set to less than 1 MHz:
The value is automatically set to 1 MHz.
2
Press
[EXIT]
to close the measurement frequency
setting screen.
3
Press
[EXIT]
to close the advanced settings
screen.
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......