46
Judging Measurement Results
3.3.2
Judging with Upper and Lower Limit Values (Comparator
Judgment Mode)
This mode judges if the measurement results are within the specified range.
The comparator judgment allows you to do the following.
• Preset a judgment reference with reference value, upper and lower limit values, and display the
judgment result as HI (higher than the upper limit value), IN (within the range set for the upper
and lower limit values), or LO (lower than the lower limit value).
• Output the judgment results to an external device (via the EXT I/O connector).
•
Judges up to four parameters with different settings.
• Beeps to notify judgment results.
Refer to “3.3.1 Setting the Judgment Mode” (p. 44).
HI
Measured value > upper limit
IN
Upper limit value ≥ measured value ≥ lower
limit value
LO
Measured value < lower limit
- - -
If reference standards have not been set
Comparator judgment order
Judgment
order
State
Judgment display
1
• When the measured value is
MEAS ERR
•
Outside the guaranteed accuracy range (Judgment for a value
outside the guaranteed accuracy range is
NOT
)
HI
2
When judging if the measured value is higher than the lower limit
value, and the judgment is Fail.
LO
3
When judging if the measured value is lower than the upper limit
value, and the judgment is Fail.
HI
4
Other than 1, 2, 3
IN
•
If measured values are outside the guaranteed accuracy range (
REF VAL
), judgment is
performed in order of judgment when the setting of
[JUDGE EXEC]
is
DO
. If
NOT
, judgment is
not performed and
HI
judgment is returned.
•
There is no comparison performed to check if the upper limit value is greater than the lower limit
value, hence an error message will not be displayed if the setting of the upper limit value and
lower limit value are reversed.
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......