![Hioki IM7580 Instruction Manual Download Page 46](http://html1.mh-extra.com/html/hioki/im7580/im7580_instruction-manual_2137306046.webp)
38
Setting Basic Settings of Measurement Conditions
3.2.6
Setting the Measurement Signal Level
Sets the Measurement signal level.
The value of the measurement signal level may change based on the sample tested.
This instrument can set the measurement signal applied to the object to be tested using the
following three methods.
Power (P) mode
Sets measurement signal level with the power (dBm) at the DUT port
50
Ω
terminal.
Voltage (V) mode
Sets measurement signal level with the voltage (V) when the DUT
port is open.
(value of dBm converted into V)
Current (I) mode
Sets measurement signal level with the current (A) when the DUT
port is in a short circuit state.
(value of dBm converted into I)
• The setting resolution of the signal level is always 0.1 dB irrespective of the setting signal
mode.
When the level is set in the voltage or current mode, input values are automatically converted
to the setting value with a resolution of 0.1 dB.
•
The measurement accuracy varies according to the measurement signal level.
See “Measurement range” (p. 274).
1
8
3
2
1
Press
[SETUP]
.
2
Press
[BASIC]
tab.
3
Press
[LEVEL]
.
4
5
4
Press
[P/V/I]
.
5
Select the signal setting mode.
[POWER]
Sets with power (dBm).
[V]
Sets with voltage (V).
[I]
Sets with current (A).
Go to the next page.
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......