288
Measurement Accuracy
12.2.3 DCR Measurement
Accuracy guaranteed
temperature and
humidity range
0°C to 40°C (32°F to 104°F), 80% RH or less (no condensation)
However, within ±5°C (±9°F) of the calibration temperature.
Calibration temperature
range
Based on the operating temperature of calibration kit.
Guaranteed accuracy
period
1 year (Ensure open/short/load calibration are performed daily before
measurement.)
Warm-up time
At least 60 minutes
Calibration face
7 mm terminal face of IM9906 Adapter (3.5 mm/7 mm) attached to the 3.5 mm
terminal of test head
(After performing open/short/load calibration with calibration kit)
Calibration kit
When products with the following specifications or equivalent are used
LOAD: 50
Ω
±0.5%
OPEN: 100 k
Ω
or more
SHORT: 10 m
Ω
or less
Measurement accuracy
± +
+
×
1
0 05
10000
100
.
Rdut
Rdut
[%]
(Number of waveforms: 128)
(Rdut: DC resistance measurement
value Unit: [
Ω
])
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......