146
Calibration
Setting the reference value
2
1
1
1
Use
/
or scroll to select the item to be
changed.
2
To change the reference value for AC
measurement:
Press
[AC DEF]
.
To change the reference value for DC
measurement:
Press
[DC DEF]
.
When the list does not display the reference
value (When the display at the top of the list is not
VAL:DEFINE
), press
[VAL
DEF]
to change the
display.
3
4
3
Press
[TO ALL No.]
.
(Only for
[AC DEF]
)
4
Select ON/OFF.
[OFF]
Sets the reference value only for the
calibration No. currently being set.
[ON]
Sets the same definition value for all
calibration points.
5
8
7
6
5
Select the reference value to be changed.
6
Set the definition value with the numeric keypad.
[-]
Inputs the unit with a minus (-) sign.
[×10
3
]
Increases the prefix of the unit.
[/10
3
]
Decreases the prefix of the unit.
[C]
Repeats the input.
[CANCEL]
Cancels the setting.
7
Press
[SET]
.
8
Press
[SET]
to close the setting screen.
What is a reference value?
The reference value is the defined value in the calibration kit, or the value of a known reference sample.
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......