Equivalent Circuit Analysis Function
4.9.5
Settings to Judge Analysis Results
You can judge whether estimation results fall within a predefined judgment standard using the comparator
function.
Setting the upper or lower limit value
You must set upper and lower limit values for the judgment standards before using the comparator
function.
1
2
3
4
1
Press
[SETUP]
.
2
Press
[CIRCUIT]
tab.
3
Press
[COMP]
.
4
Select comparator function ON/OFF.
[OFF]
Disables comparator function.
[ON]
Enables comparator function.
5
6
5
Select the constants to set judgment standard.
(1)
Press the cell corresponding to HI of any
arbitrary parameter.
(2)
Set the upper limit value with the numeric
pad* and press
[SET]
.
Settable range
-9.99999 G to 9.99999 G
* Each common numeric keypad
[-]
Inputs the unit with a minus (-) sign.
[×10
3
]
Increases the prefix of the unit.
[/10
3
]
Decreases the prefix of the unit.
[C]
Repeats the input.
[CANCEL]
Cancels the setting.
6
(1) Press the cell corresponding to LO of any
arbitrary parameter.
(2)
Set the lower limit value with the numeric
pad* and press
[SET]
.
Settable range
-9.99999 G to 9.99999 G
If the setting is such that upper limit value < the
lower limit value, the values are automatically
switched and set.
7
Press
[EXIT]
to close the setting screen.
4
Analyzer Function
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......