92
Set Measurement Conditions for Sweep Points
4.4.4
Displaying Average Values (Average)
The measured values can be averaged using the averaging function. The variations in the
displayed measured values can be reduced with this function.
•
The measured values are averaged by arithmetic averaging during analyzer measurement
irrespective of the trigger setting. (p. 41)
• When averaging is enabled, the maximum, minimum, and peak values (local maximum and
local minimum values) during search function operation use the averaged values.
1
2
1
Use
/
to enter the averaging number of times.
Settable range 1 to 256 times
[C]
Setting is turned OFF.
2
Press
[EXIT]
to close the average setting screen.
3
Press
[EXIT]
to close the advanced settings
screen.
Summary of Contents for IM7580
Page 1: ......
Page 2: ......
Page 6: ...Contents iv...
Page 8: ......
Page 18: ...10 Operating Precautions...
Page 68: ...60 Judging Measurement Results...
Page 230: ...222 External Control I O Settings...
Page 234: ...226 Saving Measurement Conditions Panel Save Function Keyboard type KEY TYPE A a a...
Page 280: ...272 Editing Data Saved in USB Flash Drive...
Page 300: ...292 Measurement Time...
Page 327: ...A15 Initial Settings Table Appx 8 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 328: ...A16 Initial Settings Table...
Page 329: ...A17 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 330: ...A18 Initial Settings Table...
Page 331: ...A19 Initial Settings Table 10 9 8 7 6 5 4 3 2 1 Appx Ind...
Page 337: ...13 09...
Page 338: ......
Page 339: ......