4.2 Setting Basic Settings of Measurement Conditions
42
Setting the measurement frequency may change the setting for the measurement signal
level.
See
"For setting range and accuracy" (p. 45)
4
Use the numeric keypad to enter the
frequency.
5
Press a unit key to confirm the setting.
6
Press
to close the setting screen.
Use
or
to select the posi-
tion of the decimal point and the unit.
• The frequency is not confirmed until a unit key is
pressed.
• The unit keys are disabled until a number is
entered.
• If you attempt to set a test frequency greater
than 5 MHz, it will automatically be reduced to 5
MHz.
• If you attempt to set a test frequency lower than
4 Hz, it will automatically be increased to 4 Hz.
See
Sets the measurement frequency to x10.
Sets the measurement frequency to x1/10.
If you make a mistake during input:
Press the
to cancel the input so far, and
start again.
Frequency Setting
Frequency Setting
Frequency Setting
Frequency Setting
Use
or
to enter each digit of the
frequency.
Holding down a digit key changes the value continuously.
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......