13.2 Measurement Range and Accuracy
395
13
Chapte
r 1
3 S
pec
ifications
The measurement accuracy is calculated from a basic accuracy, which is based on the accuracy for
impedance Z (% rdg.) and phase angle
(
°
), and the following coefficients.
Basic accuracy
Measurement conditions of basic accuracy coefficient table
•
Using the Model 9262 Test Fixture
•
Measurement speed
: SLOW2
•
Cable length
: 0 m
• Operation 60 minutes after the power is turned on.
• Open circuit compensation and short circuit compensation both being performed.
• DC adjustment ON (when DC resistance measurement)
• Temperature and humidity: 23±5
°
C, 80%RH or less
When the measurement conditions differ from the above, multiply the level coefficient (C), measure-
ment speed coefficient (D), cable length coefficient (E), DC bias coefficient (F), and temperature coeffi-
cient (G) by the basic accuracy.
The basic accuracy is calculated by determining coefficient A and B from the basic accuracy coefficient
table in accordance with the measurement frequency*1 and measurement range, and then using the
following expression.
The basic accuracy becomes the accuracy [%] of Z and accuracy [
°
] of
.
*
1
When the measurement frequency is 1.001 MHz or more, (f [MHz] +3)/4 must be multiplied to the
basic accuracy.
13.2 Measurement Range and Accuracy
Measurement accuracy = Basic accuracy × C × D × E × F × G
C: Level coefficient/ D: Measurement speed coefficient/ E: Cable length coefficient/
F:DC bias coefficient/ G: Temperature coefficient
1 k
range or more
Basic accuracy = ±
Basic accuracy = ±
100
range or less
Zx: Impedance (effective value or value obtained by the following expression) of the test sample
Zx[
] =
L [H]
(when
= 90
)
= 1 /
C [F]
(when
= -90
)
= R[
]
(when
= 0
)
10
×
Zx[
]
Range [
]
- 1
A +
B
×
- 1
A +
B
×
Range[
]
Zx[
]
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......