5.3 Normal Sweep
145
5
Chapte
r 5
ANAL
YZER Functi
o
n
Setting the start value and end value of the sweep
1.
Press
and use the numeric keypad to set the
start value of the sweep.
Settable range: 4 Hz to 5 MHz
2.
Press
,
, or
key to confirm the setting.
3.
Press
and use the numeric keypad to set the
end value of the sweep.
Settable range: 4 Hz to 5 MHz
4.
Press
,
, or
key to confirm the setting.
5.
Press
and use the numeric keypad to enter the
number of sweep points.
Settable range: 2 to 801
6.
Press
key to confirm the setting.
7.
Select the setting method for sweep points (when fre-
quency sweep).
START-STOP Settings
The sweep points are calculated linearly from
the setting values of
,
,
and .
The sweep points are calculated logarithmical-
ly from the setting values of
,
, and
.
8.
Press
to confirm the setting.
If you make a mistake during input:
press
to cancel the input and start again.
• The setting method of the sweep points can only be selected when the sweep parameter is fre-
quency and the setting method of the sweep range is
. In other cases, the setting
method of the sweep points is fixed to linear.
• If the setting method of the sweep points is changed, the horizontal axis scale of the graph display
screen changes as shown in the figures below. (The horizontal axis scale of the graph can also be
changed with the horizontal axis scale setting.)
See:
"Horizontal Axis Scale Setting" (p. 177)
Linear Setting
Log Setting
The horizontal scale is linear display.
The horizontal scale is log display.
When the
setting is
.
When the
setting is
.
Summary of Contents for IM3570
Page 2: ......
Page 8: ...Contents vi ...
Page 16: ...Operating Precautions 8 ...
Page 34: ...1 3 Screen Configuration and Operation 26 ...
Page 138: ...4 5 Setting Application Settings 130 ...
Page 160: ...5 3 Normal Sweep 152 When the setting is Procedure Analyzer Initial Screen Press 2 1 ...
Page 272: ...5 10 Equivalent Circuit Analysis Function 264 ...
Page 336: ...9 4 Deleting a Panel 328 ...
Page 410: ...13 3 About Measurement Times and Measurement Speed 402 ...
Page 418: ...14 4 Discarding the Instrument 410 ...
Page 444: ...Appendix 13 Device Compliance Statement A26 ...
Page 447: ...Index3 Index Vertical Axis Scale 183 Voltage limit 60 78 ...
Page 448: ...Index4 ...
Page 455: ......
Page 456: ......
Page 457: ......