Load Correction (Correcting Values to Match Reference Values)
Setting reference values
Enter the reference value for the parameter displayed to the left of the parameter mode for
REF1
and the reference value for the parameter displayed to the right of the parameter mode for
REF2
.
1
Touch the
REF1
key.
2
Enter the reference value with the nu-
meric keypad and touch a unit key to
accept the setting.
2
Uses a multiple of ×1.
(Touching
EXIT
without
touching a unit key will
also cause a multiple of
×1 to be used.)
Touch when you wish
to cancel input. (This
dialog box will be
closed.)
1
Settable range:
Same as the maximum display range for
the selected parameter.
(See "10.1 General Specifications"
If you make a mistake, touch the
C
key to
reenter the value.
3
Touch the
EXIT
key.
The dialog box will be closed.
4
Touch the
REF2
key and set the refer-
ence value in the same manner.
When
DC
is selected for the correction frequency setting, only reference value 1 can be set.