8.1 Setting Open Circuit Compensation
223
8
Chapter 8 Error Compens
ation
A window such as the following will be displayed if the compensation process fails. If an error message is
displayed and compensation canceled (if you touch
), open compensation will be disabled.
The open circuit compensation process is quite sensitive to noise - both noise originating externally and
induced noise.Therefore, if open circuit compensation has been interrupted with a fault, you should check
the following points before starting the compensation process again:
• Check that the test cables are properly connected.
• Check that nothing is connected to the test cables. (Open circuit compensation cannot be performed
while any test sample is connected to the test cables.)
• Check that the test leads are arranged as closely as possible to their configuration in which measure-
ment will be performed.
• During the compensation process, be sure not to disturb the test cables or to move your hand near
them.
• Execute the guarding process.
See
"Appendix2 Measurement of High Impedance Components"(p. A3)
Select
in
step 3
of [Open Compensation Setting] (p. 216) and touch
to disable the
acquired compensation data.
When Open Compensation Failed
When You Want to Make Open Circuit Compensation Data Invalid
Error
Open Compensation Setting
The compensation values that are stored internally are not cleared by the operation
described above. When ALL or SPOT is selected, the stored compensation values can be
used.
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......