92
Accuracy table
(9)
DC voltage
Measured value/MAX/MIN/AVE
Range
(automatic ranging threshold)
Accuracy guarantee range
(resolution)
Measurement
accuracy
Input impedance
(DC input)
600.0 mV
(more than 6000 counts)
0.0 mV to ±600.0 mV
(0.1 mV)
±0.5% rdg
±0.5 mV
6.7 M
Ω
±5%
6.000 V
(more than 6000 counts,
less than 540 counts)
0.000 V to ±6.000 V
(0.001 V)
±0.5% rdg
±0.003 V
6.7 M
Ω
±5%
60.00 V
(more than 6000 counts,
less than 540 counts)
0.00 V to ±60.00 V
(0.01 V)
±0.5% rdg
±0.03 V
6.1 M
Ω
±5%
600.0 V
(more than 6000 counts,
less than 540 counts)
0.0 V to ±600.0 V
(0.1 V)
±0.5% rdg
±0.3 V
6.0 M
Ω
±5%
1000 V
(less than 540 counts)
0 V to ±1000 V
(1 V)
±0.5% rdg
±3 V
6.0 M
Ω
±5%
HIOKI CM4371C961-00