![Hioki CM4141-50 Instruction Manual Download Page 72](http://html.mh-extra.com/html/hioki/cm4141-50/cm4141-50_instruction-manual_2137259072.webp)
66
Input Specifications, Measurement Specifications
Crest factor
AC current
AC Inrush
60.00 A range
600.0 A range
3 (5000 counts or less)
2.5 (more than 5000 counts,
6000 counts or less)
2000 A range
1.5 (2000 counts or less)
Frequency detection
input level
AC current
Current frequency
60.00 A range
600.0 A range
300 counts or more
2000 A range
200 counts or more
AC Inrush
Trigger level
AC Inrush
60.00 A range
Not less than +2.0 A
PEAK
or not more than −2.0 A
PEAK
600.0 A range
Not less than +10 A
PEAK
or not more than −10 A
PEAK
2000 A range
Not less than +100 A
PEAK
or not more than −100 A
PEAK
Peak detection time
width
AC current
AC Inrush
1 ms or more (with filter disabled)
(3)
Voltage measurement specifications
Overload protection
1100 V DC
1100 V AC or 2 × 10
7
V·Hz, whichever is lower
(Up to 1 minute of continuous application)
HIOKI CM4141C961-00