![Hioki CM4002 Instruction Manual Download Page 60](http://html1.mh-extra.com/html/hioki/cm4002/cm4002_instruction-manual_2137255060.webp)
52
Output function (OUTPUT)
Output rate (A-to-V conversion ratio)
Range
Output rate
Output accuracy
RMS (RMS value)
WAVE (waveform)
6.000 mA
600.0 mV / 6.000 mA
±1.0% rdg ±5 mV
(For the display count)
±3.0% rdg ±10 mV
(45 Hz to 400 Hz)
±5.0% rdg ±10 mV
( 15 Hz to 45 Hz,
400 Hz to 2 kHz)
60.00 mA
600.0 mV / 60.00 mA
600.0 mA
600.0 mV / 600.0 mA
6.000 A
600.0 mV / 6.000 A
60.00 A
600.0 mV / 60.00 A
200.0 A
200.0 mV / 200.0 A
The instrument outputs 600.0 mV as the voltage corresponding to the current range’s full-scale count of
6000
.
Only when the 200.0 A range is used, the instrument outputs 200.0 mV AC/DC as the voltage that
corresponds to the full-scale count of
2000
.
Example: When the 600.0 mA range is used, the instrument outputs 300.0 mV as the voltage that corresponds
to 300.0 mA.
Output response
RMS (RMS value)
Refresh rate: 5 times/s
WAVE (waveform)
Frequency band: 15 Hz to 15 kHz (within ±3 dB)
Continued on next page →