167
Communications Command Reference
Communications settings
Setting the auto data output function
Syntax
Command
:SYSTem:COMMunicate:DATAout <Interface>,
<Output type>
•
Interface {OFF|RS232C|USB|LAN}
Sets the interface to which data is automatically output.
•
Output type {TYPE1|TYPE2}
Sets the data output format.
Description
Specifies the interface to which data is automatically output at the completion of
measurement as well as the output type.
TYPE1
The data number, resistance value, and judgment result are output.
Normal value
x x x x x
Invalid data (not measured)
I
N V A L
Overrange
O . F .
Under range
U
. F .
Contact check FAIL
C
.
E
R R
Overheat error
H
.
E
R R
Instrument error
E
R R O R
No judgment
N
O
C
O
M P
Within the judgment reference range
P A
S S
Upper limit value < Resistance value
U
F
A
I
L
Lower limit value > Resistance value
L
F
A
I
L
Judgment not possible
U L
F
A
I
L
•
If the set upper and lower limit values are
greater than the display range’s upper
limit value, and the measured value is
[Over.F]
(overflow)
•
If the set upper and lower limit values are
less than the display range’s lower limit
value, and the measured value is
[Under.
F]
(underflow)
•
When the BDD count is at least 1 while
the comparator function and BDD
judgment are enabled
1
2 3
4
1
2 3
4
M
o h m
P A
S S
CR LF
Unit
(4 bytes)
Resistance
(5 bytes)
Terminator
(2 bytes)
Data number
(5 bytes)
Result
(6 bytes)
Blank
1
to
6 5 5
3
5
•
Returns to 1 after
the value reaches
65535.
•
Returns to 1
when AUTO
DATA OUT is set
to OFF.
* Numerical format for measured values
Range
Normal value
2 M
Ω
X
.
X X X
20 M
Ω
X X
.
X X
200 M
Ω
X X X
.
X
2000 M
Ω
X X X X
Auto
X X X X X
Numerical format for the confirmed range
TYPE2
Data is output in the format same as
:MEASure?
according to the
:MEASure:VALid?
setting.
See the
:MEASure?
query for the format.
Communications Function
HIOKI BT5525A961-00
Summary of Contents for BT5525
Page 2: ...HIOKI BT5525A961 00 ...
Page 6: ...Contents 6 HIOKI BT5525A961 00 ...
Page 48: ...48 Notifying the Judgment Result and Test Completion with a Beep HIOKI BT5525A961 00 ...
Page 76: ...76 Initializing the Instrument Reset HIOKI BT5525A961 00 ...
Page 82: ...82 Deleting Panel Data HIOKI BT5525A961 00 ...
Page 200: ...200 License Information HIOKI BT5525A961 00 ...
Page 215: ...www hioki com HIOKI BT5525A961 00 ...