27
(12) Averaging
Set and Query Measurement Averaging
Syntax
Command
:CALCulate:AVERage
<Count>
Query
:CALCulate:AVERage?
Response
<Count>
<Count> = 1 to 99 (NR1)
Only the number of impedance measurement values specified in <Count> are
averaged and output.
Example
:CALC:AVER 10
:CALC:AVER?
10
(13) Zero Adjustment
Execute Zero Adjustment and Query Result
Syntax
Query
:ADJust?
<SPOT / ALL>
Response
<0 / 1>
Description
<SPOT>=
<ALL> =
<0> =
<1> =
Executes zero adjustment for the frequency and voltage measurement in the
currently set range.
Executes zero adjustment for the currently set all frequency and voltage
measurement.
Indicates zero adjustment succeeded.
Indicates that zero adjustment has failed.
For information on zero adjustment, see the instrument instruction manual.
Example
ADJ? SPOT
0
Zero adjustment is executed in the SPOT setting and completed normally.
Clear Zero Adjustment
Syntax
Command
:ADJust:CLEar
Description
Clears zero adjustment.
Example
ADJ:CLE