6.3 Printing
90
Printing Measured Values and Decision Results ____________________
On the Measurement display, press the
key or short-circuit the PRINT pin
to the ISO_COM of the EXT I/O connector to print the measured value and deci-
sion result.
Interval Printing _______________________________________________
This function allows you to automatically print out measurement results at preset
intervals. The print interval time must be set from the Interface Selection display.
See
"6.2 Selecting the Interface" (
The setting range is 1 to 3600 seconds.
When the print interval time is set to "0", interval printing is disabled, and only
normal printing is carried out.
Operation when interval printing is selected:
1. Start printing by pressing the
key or sending the PRINT signal via
EXT I/O.
2. Elapsed time (hours/minutes/seconds) and measurement values are printed
automatically at intervals corresponding to the preset interval time.
3. Stop printing by pressing the
key or sending the PRINT signal via EXT
I/O again.
Printing Statistical Calculation Results ____________________________
From the Statistical Calculation display, press the
key to print statistical
calculation results. If no valid data exists, only the data count is printed. When
only one valid data sample exists, standard deviation of sample and process
capability indices cannot be printed.
6.3 Printing
• When using the external trigger, if you want to print after a triggered measure-
ment finishes, connect the EOM signal of the EXT I/O to the PRINT signal.
• To print all measurements continuously, connect the EOM signal to the PRINT
signal and enable the internal trigger.
• When the statistical calculation function is on and the internal trigger is
selected, the
TRIG
key or TRIG signal will trigger statistical calculation and
printing of the current measurement value.
• Valid counts are 1 to 30000. Above 30000, the count returns to 1.
• When the printed elapsed time reaches 100 hours, it resets to 00:00:00 and
continues from zero.
(Example)
After 99 hours, 59 minutes and 50 seconds:
99:59:50
After 100 hours, 2 minutes and 30 seconds:
00:02:30
• Selecting a display other than the measurement display causes interval print-
ing to stop.
HIOKI BT3564A961-02
Summary of Contents for BT3564
Page 2: ...HIOKI BT3564A961 02...
Page 44: ...3 7 Displaying Measurement Results 38 HIOKI BT3564A961 02...
Page 80: ...4 12 Reset Function 74 HIOKI BT3564A961 02...
Page 98: ...6 3 Printing 92 HIOKI BT3564A961 02...
Page 170: ...8 8 Sample Programs 164 HIOKI BT3564A961 02...
Page 182: ...10 3 Error Indication 176 HIOKI BT3564A961 02...
Page 202: ...Appendix 11 Dimensional Diagram A20 HIOKI BT3564A961 02...
Page 206: ...Index 4 HIOKI BT3564A961 02...
Page 207: ...HIOKI BT3564A961 02...
Page 208: ...HIOKI BT3564A961 02...
Page 209: ...HIOKI BT3564A961 02...
Page 210: ...HIOKI BT3564A961 02...